Chemical Indicator Classes (Types 1-6) (CRCST Domain 3: Preparation and Packaging)
AI fact-checked, not reviewed by a certified technician. Spot an error? Report it. Report an error Outline refs: Sec3 IV.F.
- Chemical indicators (CIs) show that conditions were met; a CI alone does not prove sterility. [ISO 11140-1; AAMI ST79]
- Type/Class 1 = process indicator (e.g., tape) to distinguish processed from unprocessed items. [ISO 11140-1]
- Type/Class 2 = specific test, e.g., the Bowie-Dick air-removal test. [ISO 11140-1]
- Type/Class 5 = integrating indicator; Type/Class 6 = emulating (cycle-specific) indicator. [ISO 11140-1]
Chemical indicators change color or form when exposed to one or more sterilization process variables, such as time, temperature, saturated steam or sterilant concentration. They are a fast, visible check that a package was exposed to the process and that conditions inside the package were reached. They are one part of a monitoring system that also includes physical monitors (printouts and gauges) and biological indicators.
The six types/classes
- Type/Class 1 - Process indicators. External tape and pouch indicators. They show the item went through a process so processed and unprocessed items are not confused. They do not show that the inside of the package was sterilized.
- Type/Class 2 - Specific-test indicators. Used in specific tests such as the Bowie-Dick (dynamic air removal) test.
- Type/Class 3 - Single-variable indicators. React to one critical variable.
- Type/Class 4 - Multi-variable indicators. React to two or more critical variables.
- Type/Class 5 - Integrating indicators. React to all critical variables and are designed to correlate with biological indicator performance.
- Type/Class 6 - Emulating indicators. React to all critical variables of a specific cycle; they must match the cycle they are designed for.
Internal vs external placement
An external indicator is visible on the outside of each package. An internal indicator goes inside each package, in the location the sterilant is least likely to reach, as defined by the packaging and device IFU (often the geometric center or a challenging corner of a tray). The user should check the internal CI before using the contents.
What to do if a CI fails
If an internal CI does not reach its endpoint, the package is not used. Follow policy: report it, check the load records and physical monitors, and investigate whether other items in the load are affected.
How the CRCST tests this
Expect matching questions (which class does X), placement questions (where the internal CI goes) and interpretation questions (what a color change on tape does and does not prove).
Practice questions
What does a color change on external indicator tape on a wrapped package confirm?
A. That the contents reached sterilizing conditions for the full exposure time
B. That the package was exposed to a sterilization process
C. That the biological indicator in the load will show no growth
D. That the package has been sealed and handled correctly
Answer: B
Where should an internal chemical indicator be placed in a package or tray?
A. Taped to the outside of the wrapper so it can be read before opening
B. Anywhere inside the package, since placement does not affect the result
C. In the area most difficult for the sterilant to reach, per the IFU
D. On top of the tray, where it is easiest to see when the package is opened
Answer: C
A Type/Class 5 integrating indicator is designed to:
A. React to all critical variables of the sterilization cycle
B. Show only that the package was exposed to a sterilization process
C. Test the sterilizer's air removal before the first load of the day
D. Replace biological indicators in every load, including implant loads
Answer: A
A Bowie-Dick test pack contains which type/class of indicator?
A. Type/Class 4 (multi-variable indicator)
B. Type/Class 2 (specific-test indicator)
C. Type/Class 5 (integrating indicator)
D. Type/Class 6 (emulating indicator)
Answer: B
A Type/Class 6 emulating indicator is:
A. Designed to react to one critical variable, such as temperature
B. Designed to react to all critical variables of every cycle type
C. Designed to react to all critical variables of one specific cycle
D. A self-contained biological indicator for one specific cycle type
Answer: C
Common mistakes
- Believing a changed tape indicator proves the item is sterile.
- Putting the internal CI on top of the tray where steam reaches easily.
- Using a Type/Class 6 emulating indicator in a cycle it was not designed for.
Resumen en español: Los indicadores químicos muestran que se alcanzaron ciertas condiciones, pero no prueban esterilidad por sí solos. Clase 1: de proceso (cinta). Clase 2: prueba específica (Bowie-Dick). Clase 3: una variable. Clase 4: varias variables. Clase 5: integrador, todas las variables críticas. Clase 6: emulador, específico de un ciclo. El indicador interno va en el lugar más difícil de alcanzar para el esterilizante.
Take the free 50-question readiness check
Related topics
FAQ
Can a Class 5 integrator replace a biological indicator?
For routine release of most loads, facilities may use Class 5 integrators in PCDs per AAMI guidance, but implant loads still need a BI. Follow current AAMI ST79 and facility policy.
Which indicator is in a Bowie-Dick test pack?
A Type/Class 2 specific-test indicator.